Agilent Technologies 1912A Guía de usuario Pagina 208

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 297
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 207
7 Using 8480 Series Power Sensors
182 N1911A/1912A P-Series Power Meters User’s Guide
Does this setting match the value for the sensor? (The power sensor
reference calibration factor can normally be found above the calibration
factors table on the power sensor body.)
10 To change the settings, use the and keys to highlight the Cal Fac
value field and press to display the Cal Factor pop- up. Use the
numeric keypad to enter the required value in the Cal Factor pop- up
window.
Figure 7-126Calibration factor pop-up window
11 Press
to complete the entry.
12 Connect the power sensor to the signal to be measured.
13 The corrected measurement result is displayed.
Figure 7-127Calibration factor display
%
Calibration Factor
Vista de pagina 207
1 2 ... 203 204 205 206 207 208 209 210 211 212 213 ... 296 297

Comentarios a estos manuales

Sin comentarios