Using N8480 Series Power Sensors 8
N1911A/1912A P-Series Power Meters User’s Guide 207
Use the and keys to highlight the Cal Fac value field and press
to display the Cal Factor pop- up. Use the numeric keypad to enter
97.6 in the Cal Factor pop- up window.
Press
to complete the entry.
• Connect the power sensor to the signal to be measured.
The corrected measurement result is displayed.
%
NOTE
When no sensor tables are selected and display mode is chosen, the
calibration factor used for the measurement is displayed in the upper window as shown in
Figure 8-140.