8 Using N8480 Series Power Sensors
202 N1911A/1912A P-Series Power Meters User’s Guide
Now set the frequency of the signal you want to measure. The power
meter automatically selects the appropriate calibration factor.
6 Press . On dual channel meters select the required channel.
7 Use the and keys to highlight the Frequency value field and press
to display the Frequency pop- up. Use the numeric keypad to
enter the required value in the Frequency pop- up window.
Figure 8-137Frequency pop-up
8 Confirm your choice by pressing or .
9 Press key to close the Channel Setup screen.
Now make the measurement.
10 Connect the power sensor to the signal to be measured.
The corrected measurement result is displayed.
NOTE
You can reduce the steps required to carry out the zero and calibration procedure as
follows:
1 Connect the power sensor to the POWER REF output.
2 Press and . (For dual channel meters, press
, or as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B