General Power Meter Functions 2
N1911A/1912A P-Series Power Meters User’s Guide 51
Step Detection
To reduce the filter settling time after a significant step in the measured
power the filter can be set to re- initialize upon detection of a step
increase or decrease in the measured power. Step detection can be set in
both manual and automatic measurement average modes.
Procedure
Set step detection as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to select the Step Detect setting field.
3 Press to check the step detection to on or off.
4 Press key to close the Channel Setup screen.