Chapter 2 Service
The Agilent 53210A/53220A/53230A Calibration Procedures
132 Assembly-Level Service Guide
Display softkey annunciators show:
12 Press the Cal Step softkey.
Display softkey annunciators show:
* Only displayed if Option 106 or 115 is installed.
13 Press the uWave softkey.
Display softkey annunciators show:
NOTE If your microwave source does not have an ‘output enable/disable’
feature, temporarily disconnect the cable from the microwave
input channel until after the zero-calibration is performed.
14 Press the Perform Zero-Cal softkey to perform zero-calibration.
Display momentarily shows:
**