Chapter 4 Retrofitting Options
To Retrofit the Internal Battery DC Power (Option 300)
Assembly-Level Service Guide 197
4 Slide the outside cover of the counter to the rear and remove the
cover.
Battery Assembly Installation Procedure
1 Using a medium slot-head screwdriver, gently push back the 4-pin
connector lock tab on the motherboard end of the jumper cable
that connects between the Motherboard (J1000) and the
PowerSupply board (J2). Refer to Figures 4-1 and 4-2.
Figure 4-1. Connector Locking Tab
2 Gently move the cable 4-pin connector side-to-side while pulling
upward on the cable connector until the jumper cable is free from
the motherboard.
3 Repeat steps 1 and 2 for the other end of the jumper cable where
it connects to the AC power supply assembly (6-pin connector) on
the side wall of the instrument. See Figure 4-2. Set the jumper
cable aside in a safe location in case it is needed in the future.
Tab
J1000