Agilent Technologies Agilent 8166A Manual de instrucciones Pagina 169

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 274
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 168
Signal Routing Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 169
Signal Routing
The commands in this section allow you to control Agilent 8159x Optical
Switch modules
command:
:ROUTe[n][:CHANnel[m]]
syntax: :ROUTe[n]:[CHANnel[m]]<wsp><channel_list>
description: Sets the channel route between two ports.
NOTE When you use switches with dependent connections (e.g. the 2x2 switch), it is possible that
one route configuration automatically changes another connection!
parameters: n:
m:
channel_list
the slot number of the switch module
the switch channel within the selected switch module.
e.g. for dual 1 x 2 module m = 1 for switch 1; m = 2 for switch 2
the route between left and right ports.
channel_list format: [A....Z],[1....n]
response: If an invalid route is selected the following error message is returned - "StatParamError"
example: rout3:chan1 A,2 (module in slot 3,channel 1, connect port A with port 2)
affects: All switch modules
command:
:ROUTe[n][:CHANnel[m]]?
syntax: :ROUTe[n]:[CHANnel[m]]<wsp><channel_list>
description: Queries the current channel route of the switch for a specific module and switch channel.
parameters: n:
m:
the slot number of the switch module
the switch channel within the selected switch module. Default value is 1.
response: [A.....Z],[1.....n];[A.....Z],[1.....n] as a text string.
"," separates input and output ports of a specific connection.
";" separates parallel connections (as used in 2x2 switch).
example: rout3:chan1? A,1 simple 1xN switch
rout2:chan1? A,2;B,1 (2x2 crossover switch in crossover config).
affects: All switch modules
Vista de pagina 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 273 274