Measurement Functions – The SENSe Subsystem Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 93
command:
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO?
syntax: :SENSe[n]:[CHANnel[m]]:CORREction:COLLect:ZERO?
description: Returns the status of the most recent zero command.
parameters: none
response: 0:
any other number:
zero succeeded without errors.
remote zeroing failed (the number is the error code returned from
the operation).
example: sens1:corr:coll:zero? 0<END>
affects: All power meters and return loss modules
dual sensors: Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
syntax: SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
description: Zeros the electrical offsets for all installed power meter and return loss modules.
parameters: none
response: none
example: sens:chan:corr:coll:zero:all
affects: All power meters and return loss modules
dual sensors: Command is independent of channel.
Setting parameters for the logging function sets some parameters,
including hidden parameters, for the stability and MinMax functions and
vice versa. You must use the
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing command to set
parameters before you start a logging function using the
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe command.