Agilent Technologies Agilent 8166A Manual de instrucciones Pagina 58

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 274
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 57
Instrument Setup and Status IEEE-Common Commands
58 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
*ESE?
syntax: *ESE?
description: The standard Event Status Enable query *ESE? returns the contents of the Standard Event
Status Enable Mask (see *ESE for information on this register).
parameters: none
response: The bit value for the register (a 16-bit signed integer value).
example: *ESE? 21<END>
command:
*ESR?
syntax: *ESR?
description: The standard Event Status Register query *ESR? returns the contents of the Standard
Event Status Register. The register is cleared after being read.
parameters none
response The bit value for the register (a 16-bit signed integer value):
Bit Mnemonic Decimal Value
7 (MSB) Power On 128
6Not used 0
5 Command Error 32
4 Execution Error 16
3 Device Dependent Error 8
2 Query Error 4
1Not used 0
0 (LSB) Operation Complete 1
example: *ESR? 21<END>
Vista de pagina 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 273 274

Comentarios a estos manuales

Sin comentarios