4- 24 Agilent B2900 Users Guide, Edition 2
Front Panel Reference
Config key group
Config key group
Pressing the Config key displays the following three softkeys for setting several
functions of the SMU.
Source Displays the following three softkeys for setting the SMU source operation.
Connection Channel operation and connection setup. See “Output
Connection dialog box” on page 4-25.
Filter Output filter setup. See “Output Filter dialog box” on page
4-26.
Sweep Sweep source detail setup. See “Sweep dialog box” on page
4-27.
Measure Displays the following two softkeys for setting the SMU measurement operation.
R Compen Resistance compensation ON or OFF. See “Resistance
Compensation” on page 6-16.
On a 1-channel model, pressing this key displays the ON and
OFF keys used to set the resistance compensation on or off.
On a 2-channel model, pressing this key displays the ALL, Ch
1, and Ch 2 keys. They are used to specify the channel for
setting the resistance compensation on or off.
ALL specifies channels 1 and 2.
Ch 1 specifies channel 1 only.
Ch 2 specifies channel 2 only.
Ranging Measurement range detail setup. See “Ranging dialog box” on
page 4-28.
Common Displays the following two softkeys for setting the SMU’s miscellaneous function.
Wait Source and measurement wait time setup. See “Wait Control
dialog box” on page 4-29.
Group Only on 2-channel models. Channel grouping ON or OFF.
If
this function is ON, the channels perform synchronous
channel operation. The present setting is indicated by an
asterisk on the softkey label.