2- 14 Agilent B2900 Users Guide, Edition 2
Introduction
Source/Measure Unit
Table 2-4 Limitations for using Channels 1 and 2
Table 2-5 Current Measurement Range
Channel 1 voltage V1 Channel 2 voltage V2
Current limit
a
a. I1: Channel 1 current, I2: Channel 2 current
0 < |V1| 6 V 0 < |V2| 6 V I1 + I2 4 A
6 V < |V2| 21 V I1 + I2 1.6 4 A
6 V < |V1| 21 V 0 < |V2| 6 V I1 + I2 0.625 2.5 A
6 V < |V2| 21 V I1 + I2 2.5 A
Range value Current measurement value Resolution
10 nA
a
a. Available for the B2911A and B2912A. Not available for the B2901A
and B2902A.
0 |I| 10.6 nA 10 fA
100 nA 0 |I| 106 nA 100 fA
1 A 0 |I| 1.06 A 1 pA
10 A 0 |I| 10.6 A 10 pA
100 A 0 |I| 106 A 100 pA
1 mA 0 |I| 1.06 mA 1 nA
10 mA 0 |I| 10.6 mA 10 nA
100 mA 0 |I| 106 mA 100 nA
1 A 0 |I| 1.06 A 1 A
1.5 A 0 |I| 1.53 A
3 A 0 |I| 3.06 A 10 A
10 A
b
b. Available for pulse mode. Not available for DC mode.
0 |I| 10.6 A