34 7000 Series Oscilloscopes Service Guide
2 Testing Performance
1 Set up the oscilloscope.
a Adjust the channel 1 position knob to place the baseline at
0.5 major division from the bottom of the display.
b Set the Volts/Div setting to the value in the first line in
Table 7.
Table 6 Equipment Required to Verify Voltage Measurement Accuracy
Equipment Critical Specifications Recommended
Model/Part
Oscilloscope
Calibrator
14 mV to 35 Vdc,
0.1 V resolution
Fluke 5820A
Digital multimeter Better than 0.01% accuracy Agilent 34401A
Cable BNC, Qty 2 Agilent 10503A
Shorting cap BNC Agilent 1250-0774
Adapter BNC (f) to banana (m) Agilent 1251-2277
Adapter BNC tee (m) (f) (f) Agilent 1250-0781 or
Pomona 3285
Blocking capacitor Agilent 10240-70001 +
Agilent 1250-0774 or
Agilent 11742A + Pomona
4288 + Pomona 5088A
7000 service guide.book Page 34 Thursday, October 16, 2008 12:18 PM