Agilent Technologies InfiniiVision 7000B Series Manual de servicio Pagina 55

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 118
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 54
Testing Performance 2
7000 Series Oscilloscopes Service Guide 55
1 Connect the equipment (see Figure 8).
a Use the N cable to connect the signal generator to the
power splitter input.
b Connect one output of the power splitter to the Aux Trig
input through a 50 feedthrough termination.
c Connect the power sensor to the other output of the power
splitter.
Figure 8 Setting Up Equipment for 4-Channel External Trigger Sensitiv-
ity Test
Oscilloscope
Signal
Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50 Feedthrough
7000 service guide.book Page 55 Thursday, October 16, 2008 12:18 PM
Vista de pagina 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 117 118

Comentarios a estos manuales

Sin comentarios