Agilent Technologies InfiniiVision 7000B Series Manual de servicio Pagina 43

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 118
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 42
Testing Performance 2
7000 Series Oscilloscopes Service Guide 43
15 The result from step 14 should be between +3.0 dB and -3.0
dB. Record the result in the Performance Test Record (see
page 57).
16 Move the power splitter from the channel 1 to the channel 2
input.
17 Turn off the current channel and turn on the next channel
using the channel keys.
18 Repeat steps 3 through 17 for the remaining channels, setting
the parameters of the channel being tested where
appropriate.
7000 service guide.book Page 43 Thursday, October 16, 2008 12:18 PM
Vista de pagina 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 117 118

Comentarios a estos manuales

Sin comentarios