Calibrating and Adjusting 3
7000 Series Oscilloscopes Service Guide 63
BNC(f)-to-BNC(f) (barrel connector) to the tee as shown
below.
Figure 10 User Calibration cable for 4-channel oscilloscope
3 Connect a BNC cable (40 inches maximum) from the TRIG
OUT connector on the rear panel to the BNC barrel
connector.
4 Press the Utility key, then press the Service softkey.
5 Begin the Self Cal by pressing the Start User Cal softkey.
6 When the User Cal is completed, set the rear-panel
CALIBRATION switch to PROTECTED.
Longer cable
to TRIG OUT
To Channel 3
To Channel 2
To Channel 1
To Channel 4
7000 service guide.book Page 63 Thursday, October 16, 2008 12:18 PM