Frequency specifications
[1][2]
Range Resolution Accuracy Minimum input frequency
99.999 Hz 0.001 Hz 0.02% + 5
0.5 Hz
999.99 Hz 0.01 Hz 0.005% + 5
9.9999 kHz 0.1 Hz 0.005% + 5
99.999 kHz 1 Hz 0.005% + 5
999.99 kHz 0.01 kHz 0.005% + 5
>1 MHz 0.1 kHz 0.005% + 5 @ <1 MHz
1. Overload protection: 1000 V; input signal is <20,000,000 V × Hz (product of voltage and frequency).
2. The frequency measurement is susceptible to error when measuring low-voltage, low-frequency signals. Shielding
inputs from external noise pickup is critical for minimizing measurement errors. Turning on the low pass filter may help
you to filter out the noise and achieve a stable reading.
Duty Cycle
[3]
Mode Range Accuracy at full scale
DC Coupling 99.99% 0.3 % per kHz + 0.3 %
AC Coupling 99.99% 0.3 % per kHz + 0.3 %
Pulse Width
[4]
Range Resolution Accuracy at full scale
999.99 ms 0.01 ms (duty cycle accuracy/frequency) + 0.01 ms
2000.0 ms 0.1 ms (duty cycle accuracy/frequency) + 0.1 ms
3. Notes for duty cycle specifications:
The accuracy for duty cycle and pulse width measurements is based on a 3 V square wave input to the dc 3
V range. For ac couplings, the duty cycle range can be measured within the range of 10% to 90% for signal
frequencies >20 Hz.
The range of the duty cycle is determined by the frequency of the signal: {10 μs × frequency × 100%} to {[1 – (10
μs × frequency)] × 100%}.
The pulse width (positive or negative) must be >10 μs. The range of the pulse width is determined by the
frequency of the signal.
4. Notes for pulse width specifications:
The accuracy for duty cycle and pulse width measurements is based on a 3 V square wave input to the dc 3 V range.
The pulse width (positive or negative) must be >10 μs. The range of the pulse width is determined by the
frequency of the signal.