Electrical specifications
DC specifications for U1271A and U1272A
Function Range Resolution
Accuracy ±(% of reading +
counts of least significant digit)
Test current/
Burden voltage
U1271A U1272A
Voltage
[1]
30 mV 0.001 mV - 0.05 + 20
300 mV 0.01 mV 0.05 + 5 0.05 + 5 -
3 V 0.0001 V 0.05 + 5 0.05 + 5 -
30 V 0.001 V 0.05 + 2 0.05 + 2 -
300 V 0.01 V 0.05 + 2 0.05 + 2 -
1000 V 0.1 V 0.05 + 2 0.05 + 2 -
Z
LOW
(low input impedance)
enabled, applicable for 1000 V
range and resolution only.
- 1 + 20
Resistance
[2]
30 Ω 0.001 Ω - 0.2 + 10 0.65 mA
300 Ω 0.01 Ω 0.2 + 5 0.2 + 5 0.65 mA
3 kΩ 0.0001 kΩ 0.2 + 5 0.2 + 5 65 mA
30 kΩ 0.001 kΩ 0.2 + 5 0.2 + 5 6.5 mA
300 kΩ 0.01 kΩ 0.5 + 5 0.5 + 5 0.65 mA
3 MΩ 0.0001 MΩ 0.6 + 5 0.6 + 5 93 nA//10 MΩ
30 MΩ 0.001 MΩ 1.2 + 5 1.2 + 5 93 nA//10 MΩ
100 MΩ - 2.0 +10 - 93 nA//10 MΩ
300 MΩ 0.01 MΩ -
2.0 % + 10 @ <100 MΩ
8.0 % + 10 @ >100 MΩ
93 nA//10 MΩ
300 nS 0.01 nS 1 + 10 1 + 10 93 nA//10 MΩ
Current
[3]
300 µA 0.01 µA 0.2 + 5 0.2 + 3 <0.04 V/ 100 Ω
3000 µA 0.1 µA 0.2 + 5 0.2 + 3 <0.4 V/ 100 Ω
30 mA 0.001 mA 0.2 + 5 0.2 + 3 <0.08 V/ 1 Ω
300 mA 0.01 mA 0.2 + 5 0.2 + 3 <1.00 V/ 1 Ω
3 A 0.0001 A 0.3 +10 0.3 +10 <0.1 V/ 0.01 Ω
10 A 0.001 A 0.3 +10 0.3 +10 <0.3 V/ 0.01 Ω
Diode Test
[4]
3 V 0.0001 V 0.5 + 5 0.5 + 5 Approximately 1 to 2 mA
Auto 0.0001 V - 0.5 + 5 Approximately 0.1 to 0.3 mA
1. Notes for voltage specifications:
• The accuracy of the 30 to 300 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the test leads).
• For Z
LOW
measurements, autoranging is disabled and the multimeter’s range is set to 1000 volts in the manual ranging mode.
2. Notes for resistance specifications:
• Overload protection: 1000 Vrms for short circuits with <0.3 A current.
• Maximum open voltage is <+3.3 V
• Built-in buzzer beeps when the resistance measured is less than 25 Ω ± 10 Ω. The multimeter can capture intermittent measurements longer
than 1 ms.
• The accuracy of the 300 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and thermal effect (by
shorting the test leads).
• For the ranges of 30 MΩ and 100 MΩ, the RH is specified for <60%.
• The accuracy for ranges <50 nS is specified after the Null function is used on an open test lead.
• The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 × (specified accuracy)/°C (from -20 °C to 18 °C or 28 °C to 55 °C)