Electrical specifications
DC specifications for U1271A and U1272A
Function Range Resolution
Accuracy ±(% of reading +
counts of least significant digit)
Test current/
Burden voltage
U1271A U1272A
Voltage
[1]
30 mV 0.001 mV - 0.05 + 20
300 mV 0.01 mV 0.05 + 5 0.05 + 5 -
3 V 0.0001 V 0.05 + 5 0.05 + 5 -
30 V 0.001 V 0.05 + 2 0.05 + 2 -
300 V 0.01 V 0.05 + 2 0.05 + 2 -
1000 V 0.1 V 0.05 + 2 0.05 + 2 -
Z
LOW
(low input impedance)
enabled, applicable for 1000 V
range and resolution only.
- 1 + 20
Resistance
[2]
30 Ω 0.001 Ω - 0.2 + 10 0.65 mA
300 Ω 0.01 Ω 0.2 + 5 0.2 + 5 0.65 mA
3 kΩ 0.0001 kΩ 0.2 + 5 0.2 + 5 65 mA
30 kΩ 0.001 kΩ 0.2 + 5 0.2 + 5 6.5 mA
300 kΩ 0.01 kΩ 0.5 + 5 0.5 + 5 0.65 mA
3 MΩ 0.0001 MΩ 0.6 + 5 0.6 + 5 93 nA//10 MΩ
30 MΩ 0.001 MΩ 1.2 + 5 1.2 + 5 93 nA//10 MΩ
100 MΩ - 2.0 +10 - 93 nA//10 MΩ
300 MΩ 0.01 MΩ -
2.0 % + 10 @ <100 MΩ
8.0 % + 10 @ >100 MΩ
93 nA//10 MΩ
300 nS 0.01 nS 1 + 10 1 + 10 93 nA//10 MΩ
Current
[3]
300 µA 0.01 µA 0.2 + 5 0.2 + 3 <0.04 V/ 100 Ω
3000 µA 0.1 µA 0.2 + 5 0.2 + 3 <0.4 V/ 100 Ω
30 mA 0.001 mA 0.2 + 5 0.2 + 3 <0.08 V/ 1 Ω
300 mA 0.01 mA 0.2 + 5 0.2 + 3 <1.00 V/ 1 Ω
3 A 0.0001 A 0.3 +10 0.3 +10 <0.1 V/ 0.01 Ω
10 A 0.001 A 0.3 +10 0.3 +10 <0.3 V/ 0.01 Ω
Diode Test
[4]
3 V 0.0001 V 0.5 + 5 0.5 + 5 Approximately 1 to 2 mA
Auto 0.0001 V - 0.5 + 5 Approximately 0.1 to 0.3 mA
1. Notes for voltage specifications:
The accuracy of the 30 to 300 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the test leads).
• For Z
LOW
measurements, autoranging is disabled and the multimeter’s range is set to 1000 volts in the manual ranging mode.
2. Notes for resistance specifications:
Overload protection: 1000 Vrms for short circuits with <0.3 A current.
Maximum open voltage is <+3.3 V
Built-in buzzer beeps when the resistance measured is less than 25 Ω ± 10 Ω. The multimeter can capture intermittent measurements longer
than 1 ms.
The accuracy of the 300 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and thermal effect (by
shorting the test leads).
For the ranges of 30 MΩ and 100 MΩ, the RH is specified for <60%.
The accuracy for ranges <50 nS is specified after the Null function is used on an open test lead.
The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 × (specified accuracy)/°C (from -20 °C to 18 °C or 28 °C to 55 °C)