3. Notes for current specifications:
Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse
Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse
Specification for 300 mA range: 440 mA continuous.
Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the mea-sured time before proceeding with low
current measurements.
4. Notes for diode specifications:
Overload protection: 1000 Vrms for short circuits with <0.3 A current.
Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased di-ode or semiconductor
junctions measured between 0.3 V and 0.8 V (0.3 V ≤ reading ≤ 0.8 V).
Open voltage for diode: <+3.3 V DC
Open voltage for Auto diode: <+2.5 V DC and > -1.0 V DC
AC specifications for U1271A
Function Range Resolution
Accuracy
45 Hz to
65 Hz
30 Hz to 1 kHz
1 kHz to
5 kHz
5 kHz to
20 kHz
True RMS AC
Voltage
[1]
300 mV 0.01 mV 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40
3 V 0.0001 V 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40
30 V 0.001 V 0.7 + 20 1.0 + 25 2.0 + 25 2.0 + 40
300 V 0.01 V 0.7 + 20
1.0 + 25 2.0 + 25 -
1000 V 0.1 V 0.7 + 20 1.0 + 25 - -
LPF (low pass fi lter)
enabled, applicable for
all voltage ranges and
resolution
0.7 + 20
1.0 + 25 @ <200 Hz
5.0 + 25 @ <440 Hz
--
Function Range Resolution
Accuracy
Burden voltage/Shunt
45 Hz to 2 kHz
True RMS AC
Current
[2]
300 µA 0.01 µA 0.9 + 25 <0.04 V/ 100 Ω
3000 µA 0.1 µA 0.9 + 25 <0.4 V/ 100 Ω
30 mA 0.001 mA 0.9 + 25 <0.08 V/ 1 Ω
300 mA 0.01 mA 0.9 + 25
<1.00 V/ 1 Ω
3A 0.0001 A 1.0 + 25 <0.1 V/ 0.01 Ω
10 A 0.001 A 1.0 + 25 <0.3 V/ 0.01 Ω
1. Notes for voltage specifications:
Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.
Input impedance: 10 MΩ (nominal) in parallel with <100 pF.
2. Notes for current specifications:
Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse
Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse
Specification for 300 mA range: 440 mA continuous.
Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low
current measurements.