Agilent Technologies E5500 Phase Noise Measurement System
7-25
Absolute Measurement Examples
Free-Running RF Oscillator
Defining the
Measurement
1. From the File menu, choose Open.
2. If necessary, choose the drive or directory where the file you want is
stored.
3. In the File Name box, choose “FreeRF.pnm”.
4. Click the Open button.
The appropriate measurement definition parameters for this example
have been pre-stored in this file. Table 7-8 on page 7-46 lists the
parameter data that has been entered for the Free-Running RF Source
measurement example.)
NOTE Note that the source parameters entered for step 2 in Table 7-8 may not be
appropriate for the reference source you are using. To change these values,
refer to Table 7-6 on page 7-26, then continue with step “a”. Otherwise, go
to “Beginning the Measurement” on page 7-31: