Agilent Technologies E5503A Guía de usuario Pagina 311

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 465
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 310
11-22
Agilent Technologies E5500 Phase Noise Measurement System
FM Discriminator Measurement Examples
Discriminator Measurement using FM Rate and Deviation Calibration
6. Choose the Sources tab from the Define Measurement window.
a. Enter the carrier (center) frequency of your UUT (5 MHz to
1.6 GHz). Enter the same frequency for the detector input
frequency.
7. Choose the Cal tab from the Define Measurement window.
b. Select Derive constant from FM rate and deviation as the
calibration method.
A modulation FM tone of 20 kHz and a deviation of 10 kHz is the
recommend FM rate and deviation for this procedure.
Enter the parameters into the following step.
Vista de pagina 310
1 2 ... 306 307 308 309 310 311 312 313 314 315 316 ... 464 465

Comentarios a estos manuales

Sin comentarios