9-6
Agilent Technologies E5500 Phase Noise Measurement System
Residual Measurement Examples
Amplifier Measurement Example
8. Choose the Block Diagram tab from the Define Measurement window.
a. From the Phase Shifter pull-down, select Manual.
b. From the Phase Detector pull-down, select Automatic Detector
Selection.
9. Choose the Graph tab from the Define Measurement window.
a. Enter a graph description of your choice (E5500 Residual Phase
Noise Measurement @ 1 GHz, for example).
10. When you have completed these operations, click the Close button.