9-4
Agilent Technologies E5500 Phase Noise Measurement System
Residual Measurement Examples
Amplifier Measurement Example
4. Click the Open button.
The appropriate measurement definition parameters for this example
have been pre-stored in this file. Table 9-4 on page 9-14 lists the
parameter data that has been entered for this residual phase noise
measurement example.)
5. From the Define menu, choose Measurement; then choose the Type
and Range tab from the Define Measurement window.
a. From the Measurement Type pull-down, select Residual Phase
Noise (without using phase lock loop).