Agilent Technologies E5500 Phase Noise Measurement System
14-7
Baseband Noise Measurement Examples
Baseband Noise without using a Test Set Measurement Example
4. Choose the OK button. The appropriate measurement definition
parameters for this example have been pre-stored in this file. (Table on
page 14-4) lists the parameter data that has been entered for this
measurement example.) a Test Set Measurement
Beginning the
Measurement
1. From the Measurement menu, choose New Measurement.
2. When the Perform a New Calibration and Measurement dialog box
appears, click OK.
Making the
Measurement
3. When the Connect Diagram appears on the computers display, click on
the Continue button.
Figure 14-3 Connect Diagram for the Baseband using a Test Set Measurement