2-36 Programming Guide
Programming Commands and Examples ESG Family Signal Generators
:LIST Subsystem
Load List From Step Sweep
:LIST:TYPE:LIST:INITialize:FSTep
This command eliminates the existing sweep list data and replaces it with the step sweep
data points.
Manual Point
:LIST:MANual <val>
:LIST:MANual?
This command sets the current element used by the list mode. If list mode is controlling
frequency and/or power then the indexed element in the respective list(s) will be used. If
the list mode is AUTO, then nothing will happen until the mode is changed to MANUAL. If the
point selected is beyond the length of the longest enabled list, then the point will be set to
the maximum possible point, and an error will be generated.
At *RST, this value is set to 1.
Power List
:LIST:POWer <val>{,<val>}
:LIST:POWer?
This command defines the power list. Each value specifies the power level that the
instrument uses when it is sweeping. This list is stored in a file (POW_FILE), which is only
reinitialized upon power-up preset.
Power List Points Query
:LIST:POWer:POINts?
This command queries the signal generator for the number of points in the power list.
Preset List
:LIST:TYPE:LIST:INITialize:PRESet
This command eliminates the existing sweep list data and replaces it with a new list
consisting of the following: one point at a default frequency with a default amplitude and a
default dwell time.