1-4 Programming Guide
Preparing for Use ESG Family Signal Generators
Setting up the Equipment for Remote Operation
GPIB Interconnecting Cables
The GPIB connector enables you to connect the signal generator to any other instrument
or device on the interface bus. A GPIB connector and cable are shown in Figure 1-1. The
codes next to the connector describe the GPIB electrical capabilities of the signal
generator, using IEEE Std. 488-1978 mnemonics (GPIB, IEEE-488, and IEC-625 are
electrically equivalent, although IEC-625 uses a unique connector).
Figure 1-1 GPIB Connector and Cable
Briefly, the mnemonics translate as follows:
Mnemonic Signal Description Capabilities
SH1 Source Handshake Complete
AH1 Acceptor Handshake Complete
T5 Talker Capable of basic talker, serial poll, and unaddressed if
MLA.
TE0 Talker, Extended Address None
L3 Listener Capable of basic listener, and unaddressed if MTA
LE0 Listener, Extended Address None
SR1 Service Request Complete
RL1 Remote Local Complete
PP0 Parallel Poll None
DC1 Device Clear Complete
DT1 Device Trigger Complete
C0 Controller None
E2 Identifies electrical tristate drivers