Programming Guide 3-39
ESG Family Signal Generators Remote Data Transfer
Data Transfer Troubleshooting
3. Input integers must be between 0 and 16383.
4. Each I and Q waveform must have at least 16 points (a point equals an integer).
5. Each I and Q waveform must contain an even number of points.
6. I and Q waveforms should be the same length. (Different length I and Q waveforms are
allowed. The shorter waveform will give a 0V output from its end up to the length of the
longer waveform.)