Programming Guide 2-5
ESG Family Signal Generators Programming Commands and Examples
IEEE 488.2 Common Commands
*TRG (Trigger)
*TRG
This command triggers the device if, and only if, Bus Triggering is the type of trigger event
selected. Otherwise, *TRG is ignored.
*TST? (Self-Test)
*TST?
This query returns the result of the power-up selftest:
0 - Passed (no tests failed and at least one test passed)
1 - Failed (one or more tests failed)
*WAI (Wait-to-continue)
*WAI
This command causes the instrument to wait until all pending commands are completed,
before executing any other commands.