[:CW|:FIXed]
[SOURce:]FREQuency[1][:CW|:FIXed] <frequency> selects the non-swept
sample rate for arbitrary waveforms or waveform frequency for the built-in
waveforms (sine, square, etc.).
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
frequency> numeric see below | MINimum | MAXimum Hz
MINimum selects 0 Hz.
Arbitrary Waveforms and Sine Wave Outputs: MAXimum selects the current reference
oscillator frequency divided by 4.
Square Wave Outputs: MAXimum selects the current reference oscillator frequency
divided by 16.
Ramps and Triangle Outputs: MAXimum selects the current reference oscillator
frequency divided by 4 further divided by the [SOURce:]RAMP:POINts value.
For non-sine wave outputs, multiply the MAXimum value by 2 if frequency doubling is in
effect (see the [SOURce:]FREQuency[1]:RANGe command).
The above values bound the legal range for <
frequency>.
Comments Executable when Initiated: Yes
Coupling Group: Frequency
Related Commands: TRIGger[:STARt]:SOURce,
[SOURce:]FREQuency[1]:MODE, [SOURce:]FUNCtion[:SHAPe],
[SOURce:]ROSCillator commands
*RST Condition: SOURce:FREQuency1:FIXed 10 kHz
Example Setting the Sample Rate or Waveform Frequency
FREQ 1E3 Sets the frequency to 1000 Hz.
[SOURce:]FREQuency[1]
322 Command Reference Chapter 8
Vista de pagina 323