44 Chapter 2
Basic Operation
Configuring the RF Output
Figure 2-6 Alternating Sweeps on 8757D
Configuring an Amplitude Sweep
1. Press
Return > Sweep > Off.
This turns off both the current sweep and the alternate sweep from the previous task. The current CW
settings now control the RF output.
2. Press
Configure Ramp/Step Sweep.
3. Using the
Ampl Start and Ampl Stop softkeys, set an amplitude range to be swept.
4. Press
Return > Sweep > Ampl.
The new amplitude ramp sweep settings control the RF output and the CW mode is turned off.