Chapter 2 45
Basic Operation
Configuring the RF Output
Configuring a Ramp Sweep for a Master/Slave Setup
This procedure shows you how to configure two PSGs and an 8757D to work in a master/slave setup.
1. Set up the equipment as shown in Figure 2-7. Use a 9-pin, D-subminiature, male RS-232 cable with the
pin configuration shown in Figure 2-8 on page 46 to connect the auxiliary interfaces of the two PSGs.
You can also order the cable (part number 8120-8806) from Agilent Technologies.
By connecting the master PSG’s 10 MHz reference standard to the slave PSG’s 10 MHz reference input,
the masters timebase supplies the frequency reference for both PSGs.
Figure 2-7 Master/Slave Equipment Setup
Vista de pagina 54