72 Chapter 3
Optimizing Performance
Creating and Applying User Flatness Correction
Figure 3-6 User Flatness with mm-Wave Source Module and Option 1EA Signal Generator
NOTE To ensure adequate RF amplitude at the mm-wave source module RF input when using
Option 1EA signal generators, maximum amplitude loss through the adapters and cables
connected between the signal generators RF output and the mm-wave source module’s RF
input should be less than 1.5 dB.
Configure the Signal Generator
1. Turn on the signal generators line power. At power-up, the signal generator automatically does the
following:
senses the mm-wave source module
switches the signal generators leveling mode to external/source module
sets the mm-wave source module frequency and amplitude to the source module’s preset values
displays the RF output frequency and amplitude available at the mm-wave source module output
The MMMOD indicator in the FREQUENCY area and the MM indicator in the AMPLITUDE area of the signal
generator’s display indicate that the mm-wave source module is active
NOTE For specific frequency/amplitude ranges, see the mm-wave source module specifications.