Agilent Technologies E5071C Guía de usuario Pagina 309

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 551
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 308
Chapter 13 309
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:DISP:COL{1-2}:TRAC{1-4}:DATA
Syntax :DISPlay:COLor{[1]|2}:TRAC{[1]|2|3|4}:DATA <numeric 1>,<numeric 2>,<numeric 3>
:DISPlay:COLor{[1]|2}:TRAC{[1]|2|3|4}:DATA?
Description Sets the color of the data trace of trace 1 (:TRAC1) to trace 4 (:TRAC4) for normal display
(:COL1) and inverted display (:COL2).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric 1},{numeric 2},{numeric 3}<newline><^END>
Example of use
10 OUTPUT 717;":DISP:COL1:TRAC1:DATA 1,2,3"
20 OUTPUT 717;":DISP:COL1:TRAC1:DATA?"
30 ENTER 717;A,B,C
Related commands :DISP:COL{1-2}:RES on page 308
Equivalent key
[System] - Misc Setup - Color Setup - Normal|Invert - Data Trace 1|Data Trace 2|
Data Trace 3|Data Trace 4
<numeric 1> <numeric 2> <numeric 3>
Description Amount of red Amount of green Amount of blue
Range 0 to 5 0 to 5 0 to 5
Resolution 1 1 1
Vista de pagina 308
1 2 ... 304 305 306 307 308 309 310 311 312 313 314 ... 550 551

Comentarios a estos manuales

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm