Agilent Technologies E5071C Guía de usuario Pagina 444

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 551
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 443
444 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:STAT:OPER:PTR
Syntax :STATus:OPERation:PTRansition <numeric>
:STATus:OPERation:PTRansition?
Description Sets the value of positive transition filter of the Operation Status Register.
Parameters
Note that bit 0 to bit 3, bit 6 to bit 13 and bit 15 cannot be set to 1.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":STAT:OPER:PTR 16"
20 OUTPUT 717;":STAT:OPER:PTR?"
30 ENTER 717;A
Related commands :STAT:OPER? on page 441
:STAT:OPER:NTR on page 443
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Value of the positive transition filter
Range 0 to 65535
Preset value 16432
Resolution 1
Vista de pagina 443