410 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:EXT:PORT{1-2}
Syntax :SENSe{[1]|2|3|4}:CORRection:EXTension:PORT{[1]|2} <numeric>
:SENSe{[1]|2|3|4}:CORRection:EXTension:PORT{[1]|2}?
Description For channel 1 (:SENS1) to channel 4 (:SENS4), sets the correction amount for the port
extension of port 1 (:PORT1) to port 2 (:PORT2) as the delay time.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:EXT:PORT1 1E-3"
20 OUTPUT 717;":SENS1:CORR:EXT?"
30 ENTER 717;A
Related commands :SENS{1-4}:CORR:EXT on page 409
Equivalent key
[Cal] - Port Extensions - Extension Port 1|Extension Port 2
<numeric>
Description Delay time
Range -10 to 10
Preset value 0
Unit s (second)