Agilent Technologies E5071C Guía de usuario Pagina 433

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 551
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 432
Chapter 13 433
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SOUR{1-4}:POW:CENT
Syntax :SOURce{[1]|2|3|4}:POWer:CENTer <numeric>
:SOURce{[1]|2|3|4}:POWer:CENTer?
Description Sets the center value of the sweep range for the power sweep for channel 1 (:SOUR1) to
channel 4 (:SOUR4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SOUR1:POW:CENT 0"
20 OUTPUT 717;":SOUR1:POW:CENT?"
30 ENTER 717;A
Related commands :SENS{1-4}:SWE:TYPE on page 428
:SOUR{1-4}:POW:ATT on page 432
:SOUR{1-4}:POW:SPAN on page 438
Equivalent key
[Center]
<numeric>
Description Center value
Range Varies depending on the power range.
Preset value -2.5
Unit dBm
Resolution 0.05 or 0.025