M3000A/M3046AM3015A/M3016A Service Guide
Testing the Instrument 4-21
6S2
3HUIRUPDQFH7HVW
This test checks the performance of the SpO
2
parameter. The SpO
2
performance test is
required once every two years.
7RROVUHTXLUHG none
6WHS Connect an adult SpO
2
transducer to the SpO
2
connector on the Measurement
Server.
6WHS Measure the SpO
2
value on your finger (this assumes that you are healthy).
6WHS The value should be between 95% and 100%
0DLQVWUHDP&2
3HUIRUPDQFH7HVW
This test checks the performance of the mainstream CO
2
parameter. The mainstream CO
2
performance test is required once per day and when the Instrument is repaired or when
Instrument parts are replaced.
This mainstream CO
2
performance test is a User task and is described in theUsers Guide.
There is no requirement for additional tests to be performed by the Service Engineer or
Biomedical Engineer.