M3000A/M3046A/M3015A/M3016A Service Guide
Introduction to the Instrument 1-47
3UHVVXUH6HQVRU
The pressure sensor is used to measure the ambient pressure during the auto zero process and
to measure the pressure in the measurement cell during measurement mode.
$QDORJ6HFWLRQ
The analog section amplifies and digitizes the main, reference, pressure and temperature
signals.
&RQWUROOHUDQG3HULSKHUDOV
The controller reads the digitized values from the analog section and calculates the CO
2
wave
and numerics based on the measured main, reference, pressure and temperature signals. The
controller also controls the CO
2
frontend based on the control information received from the
Measurement Server, the Optical Code Recognition and the measured cell pressure.
*DV,QOHWZLWK2SWLFDO&RGH5HFRJQLWLRQ
The gas inlet allows the connection of Microstream FilterLines. The Microstream FilteLines
are detected by the Optical Code Recognition.
6HULDO,QWHUIDFHZLWK)(/3
The Serial Interface and Frontend Link Protocol (FELP) provide the communication
interface between the CO
2
frontend and the CO
2
application software module running on the
main CPU of the Measurement Server.