M3000A/M3046A/M3015A/M3016A Service Guide
6-16 Repairing the Instrument
5HPRYLQJWKH&2
6FUXEEHU
To remove the CO
2
Scrubber, do the following:
6WHS Locate the CO
2
Scrubber in the Extension.
6WHS Being careful not to touch anything else in the Extension, use tweezers to pull the
body of the CO
2
Scrubber out of the bracket
.
6WHS Holding the body of the CO
2
Scrubber with your fingers, carefully disconnect the
Extension intake tube from the scrubber end and remove the CO
2
Scrubber from
the Extension.
6WHS Dispose of the CO
2
Scrubber according to local legal requirements for low volume
chemical waste.
5HPRYLQJWKH,QIUDUHG/DPS
To remove the Infrared Lamp, do the following:
6WHS Locate the Infrared Lamp assembly in the Extension.
6WHS Being careful not to touch anything else in the Extension, use a thin-bladed screw-
driver to pry open the locking clamp holding the Infrared Lamp assembly closed.
127( Make sure that the movement of the locking clamp does not pinch any
adjacent tubing.
6WHS Remove the Infrared Lamp cover
Vista de pagina 215