M3000A/M3046AM3015A/M3016A Service Guide
Repairing the Instrument 6-17
6WHS Holding the spring with your finger, gently pull the red lamp removal band to dis-
lodge the Infrared Lamp and the Contact Housing from the assembly.
127( Now that it is exposed, do 127 allow anything to fall into the Infrared
Lamp assembly.
127( After replacing the Infrared Lamp, reset the value displayed on the Reset
IRSourceTime selection to zero (Service Mode>CO
2
Setup).