M3000A/M3046A/M3015A/M3016A Service Guide
Introduction to the Instrument 1-23
Electrocardiogram/Respiration (ECG/Resp) Measurement
'HVFULSWLRQ
The Measurement Server has a three-channel electrocardiogram and respiration
measurement. It is designed to be used with adult, neonatal, or pediatric patients in ICU and
OR environments.
0HDVXUHPHQWV
The ECG/Resp measurement produces continuous real-time waves for both cardiac and
pulmonary activity. It also generates numerics for the average heart rate (HR), derived from
the ECG, and for the respiration rate (RR).
(&*5HVS)HDWXUHV
This illustration shows the user controls and connectors for the ECG/Resp.
Features of the ECG/Resp measurements are described in the following paragraphs.
(&*0RGHV
The QRS complexes are detected automatically.
In non-paced mode, no pace pulses are expected and no pace pulse rejection occurs. In paced
mode, pace pulses of channel 1 and 2 are annotated with a small dash on the screen.
5HVS0RGHV
In auto mode, the Monitor measures respiration and adjusts the detection level automatically.
In manual mode, the user sets the detection level for measuring respiration.
This standard 12-pin
connector will accept
ECG cable.
either a 3-lead or a 5-lead