Agilent Technologies 42A Especificaciones Pagina 143

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 168
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 142
143
Chapter 6: Replacing Assemblies
To replace the module
1 If the module consists of one card, go to step 2.
If the module consists of more than one card, connect the cables together in a
master/expander configuration. Follow the procedure "To configure a multicard
module" in chapter 2.
2 Slide the cards above the slots for the module about halfway out of the
mainframe.
3 With the probe cables facing away from the instrument, slide the module
approximately halfway into the mainframe.
4 Slide the complete module into the mainframe, but not completely in.
Each card in the instrument is firmly seated and tightened one at a time in step 6.
Vista de pagina 142
1 2 ... 138 139 140 141 142 143 144 145 146 147 148 ... 167 168

Comentarios a estos manuales

Sin comentarios