Chapter 3: Testing Performance
To Test the Single-clock, Single-edge, State Acquisition
8 Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup/hold combination selected, +0.0 ps or -50 ps.
a On the Oscilloscope, select [Define meas] Define Time - Stop edge:
falling.
b On the oscilloscope, select [Shift] Time. Select Start src: channel 1,
then select [Enter] to display the setup time (Time(1)-(2)).
c Adjust the pulse generator channel 1 Delay until the pulses are aligned
according to the setup time of the setup/hold combination selected,
+0.0 ps or -50 ps.
Disregard the Period(2) value. The settings provided in this procedure may measure
the period from rising edge to rising edge, which is not a valid measurement.
Setup Time
Data Signal
Clock Signal