Chapter 3: Testing Performance
To Test the Multiple-clock, Multiple-edge, State Acquisition
4 Configure the trigger pattern.
a Select the Trigger tab. Under the Trigger tab, select the Trigger
Functions tab.
b In the General State field, select Store nothing until pattern occurs.
Then select Replace.
c Under Trigger Sequence, locate the Label 1 = trigger pattern field.
Enter “AA” in the trigger pattern field. The trigger function should now
read Store nothing until Label 1 = AA Hex occurs then Trigger and fill
memory.