Chapter 3: Testing Performance
To Test the Single-clock, Multiple-edge, State Acquisition
2 Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup/hold combination selected, +0.0 ps or -50 ps.
a On the Oscilloscope, select [Define meas] Define Time - Stop edge:
rising.
b In the oscilloscope timebase menu, select Position. Using the
oscilloscope knob, position the falling edge of the data waveform so that
it is centered on the display.
c On the oscilloscope, select [Shift] Time. Select Start src: channel 1,
then select [Enter] to display the setup time ( Time(1)-(2)).
d Adjust the pulse generator channel 2 Delay until the pulses are aligned
according the setup time of the setup/hold combination selected, +0.0
ps or -50 ps.
Setup Time
Data Signal
Clock Signal