4-40
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
Figure 4-13 Data Processing Flow Diagram
NOTE If the analyzer has an active two-port measurement calibration, all four
S-parameters will be saved with the measurement results. All four
S-parameters may be viewed if the raw data array has been saved.
1. If you want to title the displayed measurement, refer to "Titling the Displayed
Measurement" on page 4-32.
2. Press .
3. Choose one of the following disk drives:
(If necessary, refer to the external disk setup procedure in
"Saving an Instrument State" on page 4-38.)
4. Press .
5. Define the save by selecting from the following choices:
If you select , , or , the data
Save/Recall
SELECT DISK
INTERNAL DISK
EXTERNAL DISK
Save/Recall
DEFINE DISK-SAVE
DATA ARRAY ON
RAW ARRAY ON
FORMAT ARY ON
DATA ARRAY ON
RAW ARRAY ON
FORMAT ARY ON