Making Measurements
Using Limit Lines to Test a Device
4. To define the limit as a flat line, press:
5. To terminate the flat line segment by establishing a single point limit, press:
Figure 1-59 shows the flat limit lines that you have just created with the following
parameters:
stimulus from 127 MHz to 140 MHz
upper limit of 21 dB
lower limit of 27 dB
Figure 1-59 Example Flat Limit Line
To create a limit line that tests the low side of the filter, press:
LIMIT TYPE
FLAT LINE
RETURN
ADD
STIMULUS VALUE
140 M/µ
DONE
LIMIT TYPE
SINGLE POINT
RETURN
ADD
STIMULUS VALUE
100 M/µ
UPPER LIMIT
65 x1
LOWER LIMIT
200 x1
DONE
LIMIT TYPE
FLAT LINE
RETURN
ADD
STIMULUS VALUE
123 M/µ
DONE
LIMIT TYPE
SINGLE POINT
RETURN