7-62
Operating Concepts
Modifying Calibration Kits
Specify Class Menu
Once a standard has been defined, it must be assigned to a standard "class." This is a
group of from one to seven standards that is required to calibrate for a single error term.
The standards within a single class can be assigned to the locations listed in Table 7-2
according to their standard reference numbers.
A class often consists of a single standard, but may be composed of more than one standard
if band-limited standards are used. For example, if there were two load standards—a fixed
load for low frequencies, and a sliding load for high frequencies—then that class would
have two standards.
Table 7-2 Standard Class Assignments
Calibration Kit Label: _______________________
Disk File Name: _________________________________
Class Standard Reference Numbers Standard
Class
Label 1 2 3 4 5 6 7 8
S
11
A
S
11
B
S
11
C
S
22
A
S
22
B
S
22
C
Forward Transmission
Reverse Transmission
Forward Match
Reverse Match
Response
Response and Isolation
TRL thru
TRL reflect
TRL line or match