Agilent Technologies 24A Guía de usuario Pagina 330

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 453
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 329
6-42
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Perform the 2-port Error Corrections
1. Connect adapter A3 to adapter A2 on port 2 as shown in Figure 6-14.
Figure 6-14 Two-Port Cal Set 1
2. Perform the 2-port error correction using calibration standards appropriate for the
connector type at port 1 (the connector type for adapter A1).
NOTE When using adapter removal calibration, you must save calibration sets to
the internal disk, not to internal memory.
3. Save the results to disk. Name the file "PORT1."
4. Connect adapter A3 to adapter A1 on port 1 as shown in Figure 6-15.
Figure 6-15 Two-Port Cal Set 2
5. Perform the 2-port error correction using calibration standards appropriate for the
connector type at port 2 (the connector type for adapter A2).
Vista de pagina 329
1 2 ... 325 326 327 328 329 330 331 332 333 334 335 ... 452 453

Comentarios a estos manuales

Sin comentarios