Module 5
Basic Measurement
5-33
Shield (Ground)
Guard
Force
16494A-001
(1.5 m)
16494A-002
(3.0 m)
Shield Capacitance 900 pF 1800 pF
Guard Capacitance 130 pF 240 pF
Force Resistance 160m ohm 320m ohm
Triaxial Cable for Non-Kelvin connection
Shown above are typical capacitance and series resistance of the force line. If cables are too long,
high capacitance may cause the SMU to oscillate. Keep the force-guard cable capacitance less than
600 pF. Similarly, the guard-shield capacitance must be less than 5000 pF.
Note the rather high force resistance. Kelvin connections must be used to eliminate this cable error.