Agilent Technologies B1500A Manual de servicio Pagina 86

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 110
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 85
Module 6
Low Current Measurement
6-14
Linear, with zero set to mid scale
ZERO CHECK
Display Setup
The best check is done with a LINEAR scale which brackets either side of ZERO by 100 fA or less.
A log scale is particularly unacceptable, Log plots of zero are not possible, and all readings of a log
plot must be either all positive or all negative.